Single Cell Preparation
Cell Preparation Guidelines for FACS analysis and cell sorting
Preparation of cells for flow cytometric analysis and cell sorting will vary depending on the source and specific requirements for each cells type. Two absolute requirements are that the cell suspension be single cell and that cells remain integrity at the end of preparation for all FACS analysis and cell sorting.
For a high-speed FACSAria cell sorter, additional requirements, including concentrating or enriching samples and preparing cells under sterile condition, may be required. The FACSAria is a high through-put, multi-laser, multi-parameter, cell sorter. Up to 11 fluorescence parameters (5 from 488 nm line, 3 from UV 407 line and 3 from 647 nm line) can be used to sort up to 4 populations simultaneously at rates of up to 40,000 cells/second.
Sample Preparation for cell sorting:
Samples should be concentrated optimally, depending on cell type.
For adherent cell lines and larger cells, which need the 100um tip on the sorter, sample should be concentrated close to 15 millions/mL.
For lymphocyte or other smaller cell types, which need the 35-70µm tip, sample should be concentrated close to 50 millions/mL.
70 micron cell strainer mesh (Falcon 352350)
40 micron cell strainer (Falcon 352340).
Cells should be resuspended in a low protein buffer (<2% FBS or BSA) in a Ca/Mg-free buffer (PBS), with or without DNase (20-100 µg/ml) to decrease the chance of clumping, typically electrotransporated B cells. Addition DNase to single cell suspension depends on your experimental endpoints. For instance, DNA cell cycle and apoptosis analysis need enough EDTA (2mM) in your sample buffer to inactivate any DNase. The addition of Hepes also may aid in viability by buffering the pH of the sample while sorting. Cell suspensions should be passed through a 40µm nylon mesh filter (available in flow lab) right before sorting. Samples and sort receptacles should be kept on ice (if applicable).
Sort Setup: As with all flow cytometry experiments, controls are absolutely critical. An unstained sample and each single stain controls, as well as FMO (fluorescence minus one, a control for which cells are stained with all reagents except the one of interest) controls, typically for multiple color analysis, are necessary for setting up the instrument. No controls, No sorting!!
If dim populations are needed for sorting, it may be beneficial to prepare FMO controls that allow you to visualize the true background including effects introduced by compensation.
Sorting Process: During the sort process, the sorter will display all sorting related parameters, including the number of cells sorted, the locations of sorted cells, abort rates and sort rates. A purity check is always performed after sorting by re-running a small fraction of the sorted population, followed the samples lines are rinsed with Bleach and Ethanol to prevent carryover from previous sorting sample.
Negative controls:
1. unstained cells (for each cell type examined)--This is absolutely required for each analysis (except DNA cell cycle).
2. non-specific staining control for each fluorochrome or dye and each cell type (“stain” cells with either the fluorochrome isotype control, or a secondary antibody without the primary antibody)--This indicates any non-specific binding of your antibody, and is important for your interpretation of the data.
Positive controls:
-cells stained with only one antibody-fluorochrome (or other dye) at a time (for each cell type and each fluorochrome or dye)--These are essential to accurately collect and analyze data from samples stained with more than one fluorochrome or dye.
Controls are necessary to set proper compensation.
Sample Preparation
The single most important issue for a successful sort would have to be proper sample preparation. This can be broken down into four separate components:
- Single Cell Suspensions
- Optimized Sample Concentration
- Proper Sort Buffer Recipe
- Expedient Sample Processing
Single Cell Suspensions
In order for the sorter to function properly and to be able to deliver the proper results, the cells must be in a single cell suspension, and remain that way for the duration of the sort. This becomes a more important factor when working with adherent cell lines or tissues. Achieving a single cell suspension is the goal for a perfect sample preparation. One of the easiest tricks is to remove any large aggregates is by filtration (typically through fine 30-50um nylon mesh). There are two options. The first option, and the easiest, is to use a 12x75 mm Falcon test tube with a cell strainer cap (Falcon 352235). Simply pipet the cell suspension through the top of the cap. The second option is a little more involved. To minimize sample loss, use the techniques described below. The following tools are required (sterilize as necessary):
- 1 syringe sufficient to hold sample and one additional mL (retain cap)
- 1 Luer-Lock needle (typically 18 gauge)
- 1 15mm x 15mm piece of nylon mesh (with a 30-50um mesh)
- 1 razor blade
The next step is to aspirate the sample up through the syringe needle such that your sample is loaded along with at least 1.0mL of air as a void volume.
Once the sample has been aspirated along with the 1.0mL of void volume, carefully dispose of the needle. Maintain the syringe in an upright position so that the void bubble remains at the tip of the syringe.
The next step is to sandwich the piece of nylon mesh over the tip of the syringe using the original syringe tip. The idea is to use the tip to hold the filter mesh in place. Once the tip is in place, carefully cut off the first few mm of the tip. This will allow a hole for the filtered sample to pass through. Additionally, take care to maintain the void bubble towards the tip so no sample is accidentally lost during this step.
At this point you need to invert the needle so the void bubble rises to the plunger. Now the sample is ready to be passed through the filter mesh and the void bubble should ensure no sample remains in the syringe.
Optimized Sample Concentration
Cells must be at the proper concentration in order for the sorter to function optimally. Simply put, cells that are too concentrated will have a lower recovery due to coincidence aborts (two cell that are too close together will be rejected by the machine in order to ensure purity) and cells that are too dilute will have a longer processing time (or if they are processed faster, an increased signal CV).
Ideal Cell Suspension Concentrations
Having the sample too concentrated, or too diluted can be problematic. There is no ideal concentration that works for all cell types and sort set-ups. It is a matter of understanding some of the issues and deciding what factors are most relevant to a given cell type and experimental design.
- Coincidence Aborts
While the sorter is evaluating which cell to sort, it must also determine whether it can do so it a manner that ensures the sorted material remains pure. The sorter makes this decision based on the proximity of events in time. If the desired event is too close to a potential undesired event, the machine will abort (not sort) the desired event to ensure purity. There are other modes of sorting that favor recovery over purity, so if total cell numbers are more important than purity, we can accommodate this. This should explain why we would not want the sample so concentrated that it becomes difficult to space the cells far enough apart while they are going through the sorter. If recovery is a prime concern, aborts tend to be the issue. - Signal CVs and Sensitivity
If the sample is too dilute, to get them to run at a reasonable rate can cause another set of problems. To get them to run at the appropriate rate, the sample differential must be increased. If the differential is too high, the CV's (coefficients of variation) start to become higher. This ultimately leads to less resolution and lower sensitivity. If you are trying to separate two very close populations, the CV becomes more important. - Cell Adhesion and Clumping
Adherent cells are trickier to sort than suspension cell types. Adherent cells typically like to stick to each other given the opportunity. Careful cell preparation and media can help avert this problem, but so can cell concentration. These sticky cell types typically like to be slightly more dilute to mediate clump formation. There is no absolute formula for this, it is determined by trial and error for a given cell type. The take home message is to consider running very sticky samples slightly more dilute than the below suggestions.
The following is a list of concentration ranges based on machine set-up which typically correlates with cell type:
Nozzle Size | Cell Types | Concentration (per mL) |
---|---|---|
70um | lymphocytes, thymocytes | 8-15 x 106 |
80um | activated subsets, smaller cell lines | 7-10 x 106 |
100um | larger adherent cells | 5-9 x 106 |
These guidelines are in place to suggest where to start with, but it might be easier to tend towards a little too concentrated (towards the high end of the recommended range), but bring some additional sample buffer with you so things can be diluted as necessary.
Proper Sort Buffer Recipe
This is probably one of the most important factors to achieve an ideal sort. A properly designed buffer recipe will help maintain a single cell suspension as well as keep the cells in a good physiological state. Culture media is typically a poor sort buffer (although it can be modified).
The proper design of sort buffer for both your pre-sort sample and your collected sample is crucial for a successful sort. The following will be a basic recipe and some suggestions for modifications that might be relevant to your particular experiment. Culture media is not an ideal sort buffer for two reasons: the pH regulation fails under normal atmosphere causing the media to become basic and the calcium chloride in most culture medias is not compatible with the phosphate component of the instrument sheath buffer (the Basic Sorting Buffer without additional protein) leading to precipitation of calcium phosphate crystals. Following the suggested recipes below will help maximize the recovery and viability of your sorted cells.
Basic Sorting Buffer:
- 1x Phosphate Buffered Saline (Ca/Mg++ free)
- 1mM EDTA
- 25mM HEPES pH 7.0
- 1% Fetal Bovine Serum (Heat-Inactivated)
- 0.2um filter sterilize, store a 40C
- antibiotics (pen and strep)
Optioal Collection Media:
- Fetal Bovine Serum only
or - Your own culture media with antibiotics
or - PBS if collecting cells for RNA or DNA
For Clean Lymphoid Cells
The buffer can be simplified to HBSS with 1% FBS. The additional cations in the recipe promote better viability. Since these cells are not prone to clump, the lack of EDTA is not a problem.
For Sticky Cells
Raise the concentration of the EDTA to 5mM and use FBS that has been dialyzed against Ca/Mg++ free PBS. Some activated cells become clumpy and the chelators (EDTA) help reduce cation-dependent cell to cell adhesion.
For Adherent Cells
In order achieve good single cell preparations, one must start at the moment of detaching your cells from the plate. Typically, the trypsin (or other detachment buffer) is quenched with culture media or a PBS/FBS buffer. This is problematic because it reintroduces the cations that facilitate the cells reattaching to the plate (or each other). One must use a cation-free FBS buffer in order to stop the detachment. Additionally, the level of EDTA can be increased if necessary (but too much EDTA can be deleterious).
For Samples with High Percentage of Dead Cells
If there are a large number of dead cells in the prep, it is likely that there is soluble DNA from the dead 5 cells that will come out of solution. This DNA will start to coat the cells and and lead to severe clumping. The addition of 10U/mL DNAase II to the buffer recipe will help reduce DNA associated clumpiness.
Extremely Important: After cells are sorted into collection tubes, centrifuge cells to remove diluted buffer and replenish with fresh culture media.
These suggestions should help to optimize sample preparation for both enhanced viability and enhanced recovery. It may require some more comprehensive modifications evolving from these simplistic guidelines.
Expedient Sample Processing
The sample must be prepared in as short a time as feasible to minimize stress on the cells as sorting is a relatively harsh process. Much of this can be achieved by simplifying the staining process and staggering the sample preps if more than one sample is being sorted.
Cell Physiology
Resting cells are typically very easy to sort, but most researchers have manipulated the system such that the cells are no longer in the most ideal state for processing. This can be addressed by setting up the instrumentation to run at lower pressures to minimize the stress on the cells. It is important for the researcher to convey any of these potential physiological issues to ensure the sort is properly configured.
End Point Requirements
The desired use for the sorted material can have a role in how the instrument is configured and how the sample can most efficiently be processed. Whether cells need to be viable, sterile or are used for DNA/RNA isolation can also have a role in instrument set-up.
Specimen Preparation
Samples must be filtered through (40 u mesh) before sorting. For example, you can use the following products: B-CMN-40 (by sq. feet) from Small Parts, Inc or Falcon 2235 tubes with the mesh in the caps from Becton Dickinson. The optimal concentration of material in the sample tubes (15-20) x10(6) cells /ml. Cells bigger than lymphocytes might need to be at lower concentrations.
The following measures are important to prevent sorting material from clumping:
- Concentration of fetal calf serum or BSA should not exceed 0.5%
- Use filtered buffer that is Ca2+ and Mg2+ free unless measurement depends on these ions
- Add some 0.02% EDTA
- Use anti-clumping agents, such as Accumax.(Innovative Cell Technologies, Inc., A Phoenix Flow Systems, Inc. Company)
It is highly recommended that unstained and single color stained controls be provided for optimum set-up.
Sufficient amount of collecting tubes (of 5ml or 15ml capacity) with media (up to 1ml or 3 ml per tube respectively) is necessary. To sort populations of low percentage, higher volume of media (up to 3 ml in 5ml tubes) is recommended for better cell recovery.
Media in collection tubes should contain at least antibiotics and 10% serum to increase viability of the cells. If many cells are expected to be collected, the serum concentration can be adjusted to100% for dilution with the sheath fluid. Wash all cells after sort and replenish with fresh media prior to placing back into culture or other experimental post sort design.
Cell concentration
In order to be able to take advantage of its high speed capabilities, cells for the MoFlo can be much more concentrated than for most other cell sorters. This is especially important for experiments where large numbers of cells are to be run.Cell concentration for large cell samples: 2 to 2.5 x 107 cells/ml
This allows the cells to be sorted without having to apply excessive extra pressure to the sample tube in order to achieve a rate of 20 to 25,000 events per second. However, cells at lower concentrations can also be sorted without problem, for example when a relatively small number of cells is to be sorted. We recommend resuspending the cells in PBS containing 2% FCS, but for very sensitive cells media with FCS can be used.
Cell straining / filtering
In order to minimize the possiblity of nozzle clogs, it is required that the cell sample be put through a 40 micron cell strainer before sorting. These are available from Falcon (cat. no. 2340). After filtering, the cells should be kept on ice and protected from light. For really sticky cells, it may be necessary to filter them again just prior to sorting because they can clump when sitting for longer periods of time.Sample tubes
It is highly recommended that tubes to be used as sample tubes be made out of polystyrene (clear), while those for sample collection be made out of polypropylene (opaque), because of their differing electrostatic properties. Collection tubes should be 12 x 75 mm in size, while sample tubes can be either 12 x 75 mm in size or, for large samples, 15 ml conical tubes.We recommend collecting cells into FCS or PBS containing FCS, especially when large volumes of cells will be sorted. The reason for this is because phosphate-buffered medium (i.e. sheath fluid) can cause precipitation of salts when mixed with carbonate-buffered medium (RPMI, MEM, etc.) leading to eventual problems in viability for sensitive cells. To help you plan accordingly, a very general rule of thumb is that 1 x 106 sorted cells will end up having a volume of about 1 ml after sorting. For cells which are to grow under serum-free conditions, serum-free media can be used with the collection tubes simply being changed more often. For rare cells the volume of the sorted cells won't be large enough to cause precipitation, and therefore media with serum would also be ok to use.
SORT SAMPLE PREPARATION GUIDELINES:
- Use only BD Falcon â [35]2054 or [35]2058 12X75 polystyrene tubes for control and sort samples.
Do not use other type or substitute tubes!
- Do not fill sample tubes over 2/3 full.
- Suspend cells in PBS with no more than 0.5% BSA or 0.5% FCS at up to 3 X 10 7 per mil.
- Sample cell concentration affects sort speed.
- Use the highest concentration possible while maintaining a single cell suspension.
- We can dilute samples at sorting facility but cannot concentrate them.
- Control Samples: For all sort operations a negative or unstained control and a separate single stain control sample tube for each stain to be used is required for instrument pre-sort calibration. (ie: if you are using FITC, PE, PECy7, APC, a control sample of each is required for calibration and compensation of instrument. In this case, 5 control samples total including unstained cell sample)
- GFP sort requires only a negative or not-transfected
(not-transduced) control sample.
- Control samples should be prepared the same as the sort sample from the same type of sells as experiment samples.
- Each control sample should contain a minimum of around 200,000 cells
COLLECTION TUBE PREPARATION:
- Pre-coating tubes with 20% serum or higher is recommended.
- Populations can be simultaneously be sorted into: Eppendorf tubes, any 12 X 75mm tube (BD Falcon â [35]2054 is the best ) or 17 X 120mm tube (Falcon â [35]2097)
- Tubes should be kept cold and in dark.
- Use 1ml serum per 12 X 75mm tube. (will hold about 1.5 million sorted cells).
- 3 –4 ml serum per 17 X120mm tube. (will hold about 5 million sorted cells).
SINGLE CELL SORTING:
- Single cells can be sorted at 1 – 50,000 cells into 6 to 96 well microtiter plates or onto a slide.
- Plates must be pre-filled (100 microliters of complete medium in 96 well plate).
- Keep in a sealed sandwich bag or keep a tiny chip of CO 2 (dry ice) in the plates container to keep bicarbonate-buffered pH correct.
- Can sort directly into PCR extraction buffer or other non-culture buffer.
Sample
Preparation |
|
Sample
guidelines for the cell sort
1.
Human samples must be serology tested (HIV and
Hepatitis B negative) before brought to the facility.
3.
All the samples to be used at the FCRC should be placed ONLY in the
following types of tubes:
4.
Please provide the controls - unstained and single
color controls. Preferable concentration of the controls is around 1 x
106cells/ml, volume 0.5-1.0 ml.
| |
Murine Tissue | |
Optical Platform | |
Preparation of single cell
suspensions from mouse
tissues
For collection of blood the heart must still be beating therefore mice have to be anaesthetised. If peritoneal exudate and thymus are to be collected it is best to kill mice using anaesthesia or CO2, as cervical dislocation can cause bleeding which contaminates these tissues, especially with red blood cells, making flow cytometry analysis difficult. Peripheral BloodAs the heart has to be beating for blood sampling it should be collected in accordance with home office project and personal licence procedures. If there is no requirement for recovery post-anaesthesia Euthatal (available in BRF) can be used. For collection of both plasma and
cells for analysis, whole
blood should be collected into an eppendorf tube containing 10IU/ml
heparin Isolation of plasma
Plasma can be used to assess cytokine/chemokine levels by ELISA/Cytokine Bead Assay Isolation of white blood cells (leukocytes) It is necessary to remove
the majority of the red cells from the sample to allow detection of the
white
cells especially for flow cytometry analysis of stained leukocytes.
There are a
number of ways to do this, the cheapest, fastest and most reliable is
hypotonic
shock This gives a loosely
packed pellet and the supernatant therefore has to be removed and
discarded by
aspiration using a Pasteur pipette. The sample is now ready
for red cell removal by hypotonic shock. Before starting have the
following to
hand During this process the cell
suspension should go from opaque red to clear red The volume of the pellet
is considerably reduced by this point, however, red cell ghosts remain
and have
to be removed by washing. As above, the pellet is loosely packed, so
the
supernatant has to be removed and discarded by aspiration using a
Pasteur
pipette. To remove red cell ghosts Count cells – as blood is still rich in red blood cells even after hypotonic shock and the number of white cells need to be counted, it is best to dilute the aliquot for counting in white cell counting fluid (WCF 3% acetic acid in distilled water plus a @knife point of Gentian or crystal violet). Usually 10ul into 40ul WCF gives a reasonable count on a haemocytometer. The count should be done as soon as possible after dilution as though the red blood cells are destroyed immediately, eventually the white cells will be destroyed too. Peritoneal
exudates
cells (PEC) It is impossible to recover the complete volume at any wash step and usually recovery is lowest from the first wash step Bone MarrowThis is usually isolated from the long bones – femur (hind leg) and humerus (fore leg). Marrow is extracted from the bones by flushing using a syringe and needle. It is best to use a syringe which will hold the total volume of marrow generated during flushing for the aspiration step at the end of preparation at all stages avoid foaming At this point the extracted marrow will be a mix of single cells, clumps and ‘tubes’, to produce a single cell suspension the extracted marrow should be aspirated through a series of needles of increasing gauge. The single cell suspension is now prepared SpleenThere are a number of ways to prepare a single cell suspension of spleen the quickest and simplest uses a 5ml loose fitting glass homogeniser. Count cells – as spleen is rich in red blood cells and the number of white cells need to be counted, it is best to dilute the aliquot for counting in white cell counting fluid (WCF 3% acetic acid in distilled water plus a @knife point of Gentian or crystal violet). Usually 10ul into 40ul WCF gives a reasonable count on a haemocytometer. The count should be done as soon as possible after dilution as though the red blood cells are destroyed immediately, eventually the white cells will be destroyed too. Thymus
FACS-PBSPBS supplemented with 0.1% BSA 0.1% sodium azide This is highly toxic and should not be ingested |